Automatic Kjeldahl Nitrogen Protein Analyzer
Characteristics
- Adopt microcomputer controlling process; include manual mode and automatic mode. Can be set and switched according to user’s need.
- Large array LCD display, touch key, easy to operate.
- Automatically control distill, add water, control water level are cut off water supply.
- Various security protections: digestive system security device, steam generator water shortage alarm, water level detection fault alarm.
- Stored operation procedure quantity: 250
- Instrument shell adopts specially made plastic spray steel. Working area adopts ABS anticorrosive board, Avoid chemical reagent corrosion and mechanical surface damage, acid proof and alkali proof.
- Once detected fault, controlling system will power off automatically.
Main technical parameters
- Measure variety: foodstuff,feedstuff, foods, dairy products, drink,soil, water, medicine,precipitate and chemical.
- Working mode: Automatic and manual (without titration)
- Storing data: 250 datum
- Measure range: 0.1mgN-200mgN (nitrogen content 0.1%-99%)
- Recovery rate: ≥99% (relative error, include digestion process)
- Repetition rate: relative standard deviation<±1%
- Working hours: distillation time: 5-15 minutes, the user can set automatically according todifferent samples.
- Voltage: AC 220V, 50Hz
- Cooling water consumption: distillation fraction 3L /min (water temperature is less than 20 degrees)
- Distillation fraction power supply: 1000W
Reviews